Adit S. DesaiSchool of Physics, Georgia Institute of Technology, Atlanta, Georgia 30332, USA Travis J. WilliamsNeutron Scattering Division, Oak Ridge National Laboratory, Oak Ridge, Tennessee 37831, USA Marcus DaumSchool of Physics, Georgia Institute of Technology, Atlanta, Georgia 30332, USA Gabriele SalaSpallation Neutron Source, Second Target Station, Oak Ridge National Laboratory, Oak Ridge, Tennessee 37831, USA Adam A. AczelNeutron Scattering Division, Oak Ridge National Laboratory, Oak Ridge, Tennessee 37831, USA Garrett E. GranrothNeutron Scattering Division, Oak Ridge National Laboratory, Oak Ridge, Tennessee 37831, USA Martin MourigalSchool of Physics, Georgia Institute of Technology, Atlanta, Georgia 30332, USA
(December 6, 2023)
Abstract
A modern cold triple-axis spectrometer is planned for the High Flux Isotope Reactor (HFIR) at Oak Ridge National Laboratory. Here, we describe the conceptual principles and design of a secondary spectrometer using a multiplexed, prismatic analyzer system relying on graphite crystals and inspired by the successful implementation of the Continuous Angle Multiple Energy Analysis (CAMEA) spectrometers at the Paul Scherrer Institute [1]. This project is currently known as MANTA for Multi-analyzer Neutron Triple-Axis. We report Monte-Carlo ray-tracing simulations on a simple but realistic sample scattering kernel to further illustrate the prismatic analyzer concept’s workings, calibration, and performance.Then, we introduce a new statistical analysis approach based on the prismatic analyzer concept to improve the number of final energies measured on the spectrometer. We also study possible changes in the CAMEA design for MANTA to deliver a world-class triple-axis instrument for neutron spectroscopy of quantum condensed matter systems.
I Introduction
The neutron triple-axis spectrometer (TAS) is an important tool for studying collective excitations of condensed matter systems [2]. In a TAS experiment, the dynamical scattering function of the sample, , is measured (in selected regions of momentum-energy space) according to the scattering relations for momentum and energy transfer [3]. In simplified terms, this requires knowledge of the neutrons’ initial and final wave-vectors , the scattering angle of the detector, and the azimuthal orientation angle of the sample in the -scattering plane, which are related by
(1) |
(2) |
In the traditional TAS technique, is measured in a point-by-point fashion (within a resolution volume). Measuring the entire sample’s response in a given scattering plane typically requires a large number of combinations of (), (), , and , resulting in long measurement times when a survey of -space is desirable. Multiplexing techniques can alleviate this bottleneck by using several analyzers to measure multiple -points simultaneously. Although this can complicate the scattering geometry, the spectrometer’s resolution function, and the available background mitigation techniques, this can be advantageous over neutron time-of-flight (TOF) spectroscopy for constrained sample environments, when combined with polarization analysis techniques, to maximize neutron flux using focusing optics, and/or when only a single scattering plane is needed.
Following the distinction made previously [1], multiplexing spectrometers can be broadly categorized into two families: wide-angle multiplexing and local multiplexing. In wide-angle multiplexing – also described as multi-channel detector systems – several analyzer channels are placed around the sample in a concentric arc, generally with fixed (). This technique increases the solid angle coverage of the analyzer system but restricts the geometry. One of the earliest examples of this concept is the MADBox spectrometer[4] at the Forschungsreaktor München (FRM), which made use of 61 in-plane analyzer channels. The FlatCone design at the Institut Laue-Langevin (ILL)[5] built upon this concept and uses 31 analyzer channels placed around the sample which scatter out of plane. A particularly efficient example of wide-angle multiplexing is the MACS spectrometer at the National Institute of Standards and Technology (NIST)[6], which uses analyzers with variable and a combination of filters to suppress higher-order wavelengths. The local multiplexing technique places multiple analyzers on the same channel. An example is the IMPS [7] spectrometer, which places several analyzers at slightly offset angles from each other. The distinct analyzer crystals or blades direct neutrons toward a different spot on a position-sensitive detector (PSD). Similar concepts have been utilized in the SIKA, RITA-II, and PUMA spectrometers [8, 9, 10]. A particularly novel design is the HODACA concept, which uses a Rowland ”anti-focusing” effect by placing analyzers on a circumferential arc, each of which focuses towards a different detector, result a constant but large coverage [11].
One of the more recent developments in multiplexing TAS design is the Continuous Angle Multiple Energy Analysis (CAMEA) spectrometer, which is installed and commissioned at the Paul Scherrer Institute (PSI) [1, 12, 13]. CAMEA uses the nearly perfect transmission of highly-oriented pyrolytic graphite (HOPG) crystals [14] to combine the local and wide-angle multiplexing techniques. While traditional local multiplexing (e.g. IMPS [7]) requires analyzers to be placed at slightly different scattering angles, CAMEA places analyzers directly behind each other, simplifying the geometry. Each of the eight (8) arrays of analyzers, or stations, focuses a band of different () out of the scattering plane at specific locations on a position-sensitive detector, hence creating a quasi-energy-resolved detector. Eight (8) channels of this multi-analyzer system are placed at different angles, resulting in wide-angle coverage. A similar concept has been successfully implemented at spectrometers like MultiFlexx [15] and BAMBUS [16] and it is planned for BIFROST at the European Spallation Source (ESS) [17].
Maintenance and upgrade plans for the High Flux Isotope Reactor (HFIR) at Oak Ridge National Laboratory[18] offer the opportunity to design and execute the construction of a state-of-the-art cold-neutron triple-axis spectrometer on a new cold guide-hall location, NB-6. In this paper, we employ Monte-Carlo ray tracing simulations to explore the concept and the performance of a secondary spectrometer relying on the CAMEA design and an optimized primary spectrometer. The design principles, optimization, and performance of the primary spectrometer (guide system, velocity selector, and monochromator) are discussed in Ref.[19]. The secondary spectrometer is currently known as the Multi-Analyzer Neutron Triple Axis (MANTA) spectrometer. This project offers the opportunity to explore and implement various forms of prismatic analysis strategies, which greatly enhances the resolving power of the spectrometer, as shown conceptually in Fig.1.
This paper is organized as follows. In Sec.II, we outline the geometrical and technical parameters of our model instrument, as well as the multiplexing and prismatic analysis concepts it utilizes. In Sec.III, we detail the methods of our ray-tracing Monte-Carlo simulations on a realistic sample kernel to benchmark the performance of our model instrument. In Sec.IV we analyze the simulation data and present the principles of prismatic analysis along with the demonstration of a novel approach called Positionally-Calibrated Prismatic Analysis (PCPA). In Sec.V, we present the results of our ray-tracing simulations and data analysis approach on a realistic sample kernel and present how modifications to the original design impact data acquisition and quality. We conclude this work in Sec.VI.
II Instrument Concept and Model
The model of the primary spectrometer comprises an optimized neutron guide system with supermirror coatings that directs neutrons from the HFIR cold-neutron source (with the parameters of the existing HFIR cold-source) onto a 300mm 168mm doubly focusing monochromator, see Ref.[19] for details. The sample is located 160cm from the monochromator. An aspirational engineering rendering of the entire spectrometer in the HFIR cold guide hall is presented in Fig.2(a). Using a focusing monochromator increases the neutron flux on the sample. However, it yields a broad divergence of the incident beam at the sample position, affecting the spectrometer’s resolution. By incorporating the guide system and the doubly focusing monochromator in our simulations (see Section III), these resolution effects can be accounted for. In some instances in SectionV, we utilized a simplified version of the full primary spectrometer with an idealized, nearly monochromatic virtual source and a curved monochromator.
For the secondary spectrometer, we adapted the back-end design of CAMEA [1] to our virtual instrument, see Fig.2(b) for an engineering diagram. The design consists of angular channels, each spanning , see Fig.2(b). Each channel comprises analyzer stations in series, each built from blades of flat, highly-oriented pyrolytic graphite (HOPG) crystals. The analyzer blades of a given station have a mosaic of 1, equivalent to 60 arc minutes (60’), and are rotated to fulfill the Bragg condition for a given . We label the analyzer station with index . Furthermore, analyzer stations between different channels are focused using Rowland geometry [3, 20]. Analyzer stations placed further from the sample are larger to account for decreased solid angle coverage for a given crystal size with increasing distance. Additional details of the instrument model are provided in Tab. 1. In some instances in SectionV, we modify the CAMEA design by changing the number of stations and the analyzer mosaic.
Analyzer Station #, | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 |
---|---|---|---|---|---|---|---|---|
Energy (meV) | 3.21 | 3.38 | 3.58 | 3.8 | 4.05 | 4.33 | 4.64 | 5.01 |
Sample to Analyzer Distance (mm) | 939 | 994 | 1057 | 1120 | 1183 | 1246 | 1310 | 1375 |
Analyzer to Detector Distance (mm) | 707 | 702 | 700 | 701 | 703 | 709 | 717 | 727 |
HOPG Segment Length (mm) | 72.0 | 82.0 | 92.0 | 103 | 113 | 120 | 129 | 140 |
HOPG Segment Height (mm) | 12.0 | 12.0 | 12.5 | 13 | 13.5 | 14 | 15 | 16 |
The analyzer blades reflect neutrons from the sample to an out-of-scattering plane direction and towards a series of horizontal position-sensitive detectors (PSDs), laid out in a radial pattern above the stations. Each detector has an active length of m. By focusing neutrons on specific locations on the detectors, each analyzer station generates a “strip” of neutrons associated with a particular . As illustrated in Fig.3 for the parameters of Tab. 1, this approach achieves local multiplexing by avoiding overlap between strips generated by different stations. Thus, at minima, this allows the acquisition of values per channel.
Though this method of discrete angular and analyser channels to form a multiplexing instrument as a viable approach for decades, one of the advantages and innovations of the CAMEA concept is the utilization of the prismatic analyzer concept[21, 1, 12]. The foundations of prismatic analysis come from a first-order expansion of Bragg’s law from an analyzer crystal (with lattice spacing ). This yields a linear relationship between small deviations in incident angles and small deviations in reflected neutron wavelength around a nominal Bragg reflection (), as illustrated in Fig.4(a). As a result of analyzer mosaic, neutrons with proximate wavelengths from the nominal wavelength are Bragg-scattered in different directions, Fig.4(b). Using a position-sensitive detector, this effect can be advantageously used to encode small wavelength deviations into detector positions, Fig.4(c). Since its inception, prismatic analysis has been featured in the design of the planned time-of-flight instruments BIFROST [17] and FARO [22].
III Simulations: Methods and Sample Kernels
To assess the performance and validate the concept of our spectrometer model, we employ neutron ray-tracing simulation using the McStas package[23]. McStas is a state-of-the-art tool used in designing modern neutron scattering instruments [17, 22, 16, 7, 6, 15, 1, 9, 24, 25, 26, 27, 28]. The program uses Monte Carlo techniques to simulate the behavior of neutron scattering instrument components with high accuracy. It also allows testing the instrument’s performance on various simple sample scattering kernels to mimic actual experiments. In this work, we employed McStas versions 2.7 and 3.1.
The primary spectrometer, from the cold source to 15cm before the sample, was simulated in McStas for various choices of incident energy , using the design of Ref.[19]. These results were pre-compiled to form a set of simulated incoming beam profiles for the secondary spectrometer. The latter was simulated by adapting the McStas code of the PSI-CAMEA back-end, which was generously provided to us by its designers [1]. The virtual instrument comprises analyzer blades per station, stations per channel, and channels (for a total continuous scattering angle coverage of ). While the actual spectrometer uses filters and a radial collimator after the sample, we did not include these components in our simulations to maintain a continuous coverage; including collimators in our simulations is left for future work. A 10 10cm rectangular slit simulates the opening in front of the analyzer stations, allowing full illumination of the analyzers and unrestricted angular coverage. Unless otherwise noted, we used an infinitely thin HOPG analyzer with a mosaic of 60’. Note that modeling the analyzer in this manner does not take into account the background signal that arises from thermal diffuse scattering and phonons in HOPG, nor the effect of the decrease in peak reflectivity as a function of crystal mosaic [14, 29, 3]
We employed two types of sample kernels. The first virtual sample was a full cylinder of polycrystalline vanadium with a 2cm height and a 1.5cm radius. We used this incoherent scattering kernel to obtain calibration runs (see Section IV) of the analyzer system and the position-sensitive detectors. Then, a virtual single crystal with a similar cylindrical geometry was prepared using the McStas phonon kernel component to mimic a simple bosonic excitation (acoustic magnon or phonon) dispersing in three dimensions with a bandwidth below 2meV and uniform intensity across the Brillouin zone. The parameters of this inelastic scattering kernel are found in Tab.2.
Sample Kernel Parameters | |
radius (cm) | 1.0 |
height (cm) | 2.0 |
(barns) | 0.171 |
(barns) | 0.32 |
fcc Lattice Constant () | 6.28 |
Scattering Length (fm) | 9.405 |
Mass (amu) | 207.2 |
Speed of Sound | 2.5 |
Temperature (K) | 10 |
Debye-Waller Factor | 1 |
Simulations were conducted with the single crystal kernel to model a realistic scattering experiment covering energy transfers between meV and meV. As the detector system has an overall coverage of , the single crystal sample was measured for several positions of the detector system. Using the lowest angle of the detector systems as a reference, we measured with values around (3 positions to account for the gap between the detectors), (5 positions), and (4 positions), for a net total of 12 positions. For each detector position, the single crystal sample was rotated in steps of and we ran the Monte-Carlo process with neutrons for each sample orientation. This was repeated for two values, 5.1 meV and 5.27 meV, to produce the simulated results. The kernel was designed to measure the entire dispersion curve with meV, and the second incident energy of meV was selected to fill in the gaps in energy space. These results are shown in Section V.
IV Prismatic Analysis and Data Processing
For a given analyzer channel and orientation of the sample, it is possible to use prismatic analysis to obtain the scattering function for significantly more values of energies than the number of stations as shown conceptually on Fig.1. While prismatic analysis relies on post-measurement data analysis techniques that require careful calibration, it is possible to devise several such strategies.
CAMEA has successfully implemented an “energy-calibration” approach [30, 12, 13]. As shown in Fig.3, the neutrons scattered by a given analyzer station – nominally associated with – produce a well-separated strip (or analyzer-specific area) on the position-sensitive detector. Each strip in the detector space is broken into prismatic sub-regions (with ranging from to ). The integrated signal detected in a given sub-region is entirely associated with a particular , proximate, but not equal to the reference , which effectively yields a continuous energy coverage but requires careful calibration[13]. To determine the average analyzed by the -th detector sub-region associated with analyzer station , a calibration is performed by measuring the scattering from a vanadium sample as a function of small steps in ( meV). The incoherent vanadium signal integrated over a given sub-region yields a Gaussian profile as a function of energy, which can be fitted to obtain , hence the name of Energy-Calibrated Prismatic Analysis (ECPA). Because the efficiency of the prismatic analysis across sub-regions is not uniform, this approach also conveniently yields an intensity normalization to take analyzer transmission and prismatic/detector efficiency into account.
While the above approach works very well, it is interesting to compare it to an alternative method of data analysis which we dub the Positionally-Calibrated Prismatic Analysis (PCPA) technique. As opposed to defining discrete sub-regions in detector space and calibrating the average energy being detected, the alternative approach defines discrete steps for the final neutron energies and reconstructs the spatial distribution of each on the detector system. This approach is demonstrated in Fig.5. The rationale for this approach is that neutron detection along the detector is very well-defined by the detector electronics, which for the CAMEA implementation at PSI has 1024 pixels over an active length of m. Just as with ECPA, calibration with a vanadium standard is essential.
To calibrate the analyzer-detector system, the primary spectrometer is tuned to a particular incident energy to analyze the elastic incoherent scattering from the vanadium kernel. We have verified that the -position of the secondary spectrometer does not affect the results; thus, the calibration procedure is performed at a single position of the detector system. After scattering from the vanadium and a given analyzer blade, neutrons with energy are detected at particular locations on the PSD. This yields a Gaussian distribution for each , as seen in Fig.5. This process is repeated by sweeping across various bands of energy centered around the nominal of the analyzers, that is meV is scanned in steps of meV for (total of ’s per station ).
This process provides the distribution of neutron counts versus detector position for each discrete within the calibration set. When comparing the different ’s, it is clear that the total detector-integrated intensities vary significantly due to the analyzer’s mosaic distribution and the radial arrangement of the detectors, see Fig.5. To correct for this effect, we introduce a scale factor, , where is the integrated intensity of a given over the entire detector, therefore summing over all Gaussian profiles ascribed to a specific . The integrated intensity of the meV energy was arbitrarily chosen as a reference as it is the most intense. The scale factor is then applied as a prefactor to each Gaussian profile in the detector space. The last, and central step is to transpose the ensemble of these scaled distributions (of neutron counts versus position at fixed ) into a new set of distributions representing neutron counts versus at fixed position. In other words, we create normalized histograms of final energies for each pixel, representing the likelihood that a neutron detected on a given pixel had energy . A representative result of such a histogram is shown in Fig. 6. Note that the scaling factor only affects the relative distribution of energies within a histogram; in the end, the scaling does not affect the net flux at the detector. In the next section, we employ the PCPA technique on a realistic sample kernel for a virtual spectrometer to asses its feasability.
V Results and Secondary Spectrometer Optimization
V.1 Original Model
We now turn to our results of the expected performance of the full MANTA model discussed in Sec.II using the single-crystal kernel and the simulations methods of Sec.III and the PCPA calibration technique discussed in Sec.IV. In Fig.7, we present the results for a complete set of single-crystal rotations as constant energy-slices representing the neutron scattering intensity in the scattering plane for three different energy transfers spanning the bottom, middle, and top of the dispersion band. In Fig.8, we present momentum-energy slices across the same datasets. These results highlight the efficiency of the secondary spectrometer in capturing the entire volume of with only two incident energies . Furthermore, it reveals resolution effects that are expected on triple-axis spectrometers, namely focusing and de-focusing effects, see e.g. the distribution of intensity on excitation rings in Fig.7(a) and Fig.7(b), as well as energy-broadening due to momentum integration and analyzer mosaic, see e.g. Fig8(b).
V.2 Studies of Analyzer Mosaic Effects
By its nature, prismatic analysis, particularly the PCPA technique we introduce, invites exploring changes to the original instrument model to measure more final energies simultaneously. In the following two subsections, we study two possible modifications of the instrument hardware: (1) changing the analyzer mosaic and (2) changing the number of analyzer stations. Analyzer crystal mosaic is an important parameter in determining the resolution of a triple-axis spectrometer: the larger the crystal mosaic, the larger the bandwidth of energies scattered by the analyzer. In a traditional setup, analyzers with a large mosaic () result in a poorer energy resolution. However, the foundational principle of the prismatic analysis is that the energy resolution can be independent of crystal mosaic through the use of distance collimation [21, 31, 1].
To test this idea, we modified our virtual model to coarsen the analyzer mosaic to instead of . The broader bandwidth of energies scattered by a single analyzer imply that the secondary spectrometer measures final energies more hom*ogeneously, as seen from our calibration runs on Fig.9(a). As detailed in Sec.IV, to produce pixel-energy histograms in Fig.9(b), we first re-scale the profiles of Fig.9(a) to account for intrinsically weaker scattering of specific energies. However, the scale factor will also correspondingly scale the signal uncertainty. Increased analyzer mosaic lowers the scale factor for most incident energies , reducing the statistical error from scaling the Gaussian distribution. We note that several effects not accounted for in our simulations may limit the applicability of this concept. A mosaic will require thicker crystals to account for the decrease in peak reflectivity, which will result in larger background effects from thermal diffuse scattering and inelastic phonon scattering from the crystals themselves [14, 29]. This effect is complex and has not been modeled in our McStas simulations. The effect of and mosaic for our sample kernel simulations is shown in Fig.10.
V.3 Towards Continuous Energy Coverage
With 8 analyzer stations, there is a gradual increase in the spacing of the analyzers, resulting in a non-continuous coverage of energy transfer. This increase in spacing with increasing is chosen due to the coarser energy resolution for the larger analyzers, thereby minimizing overlap between the energies scattered by differing stations. To eliminate gaps in the measured neutron scattering spectrum, an additional measurement, using, for instance, meV, must be combined with the original , which doubles the measurement time. However, with the PCPA method, it is, in principle, ideal to include overlap in the energies an analyzer will measure. The calibration accounts for, and in fact improves if multiple analyzers scatter neutrons with the same energy . This is because the uncertainty of using the scale factor for normalization is reduced.
Using this guiding principle, we modified the virtual secondary spectrometer to include 10 analyzer stations, using constant steps in the analyzer’s energy , beginning at 3.20 meV and increasing in intervals of 0.20 meV up to meV. The upper limit of meV is constrained by the need to use a beryllium filter to remove higher-order neutron wavelength contamination from the scattered beam, while the lower limit of meV was chosen due to increasingly large scattering angles required for low-energy measurements. To explore a cost-effective alternative, we also created a 5 stations design with even intervals of 0.40meV over the same range. By performing simulations with and meV, it is possible to obtain a complete neutron scattering spectrum using the PCPA method.
To compare all three designs at different analyzer mosaic values and reduce computation time, we conducted a similar simulation procedure described in Section III but with a few modifications. First, an idealized, nearly monochromatic virtual source within McStas and a curved monochromator created a beam directed at the vanadium kernel for calibration and simulation of the single crystal kernel. This simplified primary spectrometer was less computationally costly than a full simulation. We repeated the vanadium calibration process in a nearly identical manner for each design. However, for large analyzer mosaics, final energies that were previously negligible need to be accounted for, and we used additional calibration energies to properly interpolate the signal. Finally, all designs used a second in addition to . The 5 station design used , the 8 station design again used , and the 10 station design used . The results, shown in Fig.10 for a typical momentum-energy slice, show that the integrated count rate increases with mosaic with little visible degradation in resolution. Note, however, that these simulations do not consider the decrease in peak reflectivity, and therefore the increase in total count rate will be partially mitigated. As mentioned previously, the increase in mosaic is not a fully independent parameter and will result in other effects not fully modeled in McStas, such as background from thermal diffuse scattering.
To better quantify the quality of the results, we investigated the single crystal dispersion signal’s energy broadening due to MANTA’s finite energy resolution. We did this by fitting the data in Figs.8(d) and Fig.10 to constant- cuts separated by intervals of 0.1 and subsequently fitted with a Gaussian energy profile using bins with a width of 0.04meV. The subsequent fitted Gaussian FWHMs as a function of are shown in Fig.11. The effective energy broadening on the dispersive signal ranges from 0.2 to 0.3meV, somewhat coarser than for a cold triple-axis spectrometer, an effect we attribute to our virtual experiments’ open, uncollimated geometry and the integration over a sloped, dispersing signal. Additionally, Fig.11 shows that the energy broadening between the mosaic simplified model and the mosaic simplified model is nearly identical, but taking into account the full primary spectrometer coarsens the energy resolution to up to 0.4meV. Using prismatic analysis, the dispersion energy resolution appears independent of crystal mosaic, which agrees with the results reported in the original prismatic concept paper [21]. It is also worth noting that despite the potential background challenges from coarse mosaicity, using a mosaic as high as is not without precedent. The indirect time-of-flight spectrometer FARO [22], which also uses prismatic analyzers, decided to use this coarse mosaic value.
VI Conclusion
In this work, we have performed Monte-Carlo ray-tracing simulations of several multiplexed triple-axis spectrometer models relying on prismatic analysis to obtain a quasi-continuous energy-transfer coverage with only one or two incident energies . Building on the concept and the design of the CAMEA spectrometer at PSI, we have introduced a distinct data analysis approach relying on the statistical likelihood that a neutron detected on a given pixel corresponds to a particular final energy . This approach is dubbed Positionally-Calibrated Prismatic Analysis (PCPA). Simulations of the spectrometer model for a simple inelastic scattering signal show that PCPA data analysis maximizes the number of final energies measured simultaneously. Departing from the CAMEA design with eight analyzer stations, we explored further design evolutions, including changing the number of analyzer stations and the crystal mosaic of the graphite analyzers. These conceptual studies highlight the enormous flexibility offered by prismatic analysis. In principle, any existing TAS with a PSD located far enough from the analyzer to allow for distance collimation effects to occur can make use of the prismatic analyzer concept and the PCPA technique. Our conceptual work serves as a foundation for performing more comprehensive studies of multiplex prismatic spectrometers, taking collimation and background mitigation into account. This includes testing a more complex sample kernel, considering thermal diffuse scattering from the analyzers, and dealing with complex sample environments. Our work utilized a realistic primary spectrometer developed in Ref.[19], including a fully optimized neutron guide system for the cold source of the High Flux Isotope Reactor. Although no formal concept has been chosen for the secondary spectrometer, when combined with a refined, CAMEA-style multiplexed prismatic system, a concept known as MANTA, our work shows that this modern beam-line would deliver world-class performance to study excitations in quantum condensed matter physics.
Acknowledgements.
The work at Georgia Tech was supported by the Department of Energy, Basic Energy Sciences, Neutron Scattering Program under grant DE-SC-0018660. Initial work by A.D. was supported by the Georgia Institute of Technology’s Letson Fellowship and the President’s Undergraduate Research Award. We are grateful for the fruitful discussions with Mark Lumsden, Lowell Crow Jr., Jaime Fernandez-Baca, Barry L. Winn, Michael Hoffmann, Ian Turnbull, Mads Bertleson, Daniel Mazzone, and Jonas Birk that helped advance this work. We are indebted to Felix Groitl for providing is with the McStas model of CAMEA.
References
- Groitletal. [2016]F.Groitl, D.Graf,J.O.Birk, M.Markó, M.Bartkowiak, U.Filges, C.Niedermayer, C.Rüegg,andH.M.Rønnow,CAMEA–a novel multiplexing analyzer for neutron spectroscopy,Review of Scientific Instruments87,035109 (2016).
- Brockhouse [1995]B.N.Brockhouse,Slow neutronspectroscopy and the grand atlas of the physical world,Reviews of Modern Physics67,735 (1995).
- Shiraneetal. [2002]G.Shirane, S.M.Shapiro,andJ.M.Tranquada,Neutron Scatteringwith a Triple-Axis Spectrometer: Basic Techniques(Cambridge University Press,2002).
- Demmeletal. [1998]F.Demmel, A.Fleischmann,andW.Gläser,A multi-detector systemfor a triple-axis spectrometer,Nuclear Instruments and Methods in PhysicsResearch Section A: Accelerators, Spectrometers, Detectors and AssociatedEquipment416,115(1998).
- Kempaetal. [2006]M.Kempa, B.Janousova,J.Saroun, P.Flores, M.Boehm, F.Demmel,andJ.Kulda,TheFlatCone multianalyzer setup for ILL’s three-axis spectrometers,Physica B: Condensed Matter385-386,1080 (2006).
- Rodriguezetal. [2008]J.A.Rodriguez, D.M.Adler, P.C.Brand,C.Broholm, J.C.Cook, C.Brocker, R.Hammond, Z.Huang, P.Hundertmark, J.W.Lynn, N.C.Maliszewskyj, J.Moyer,J.Orndorff, D.Pierce, T.D.Pike, G.Scharfstein, S.A.Smee,andR.Vilaseca,MACS—a new high intensity cold neutron spectrometer atNIST,Measurement Science and Technology19,034023 (2008).
- Jiménez-Ruizetal. [2006]M.Jiménez-Ruiz, A.Hiess, R.Currat,J.Kulda,andF.Bermejo,IMPS: A multianalyser detector system for the thermalthree-axis spectrometer IN8,Physica B: Condensed Matter385-386,1086 (2006).
- Wuetal. [2016]C.-M.Wu, G.Deng, J.Gardner, P.Vorderwisch, W.-H.Li, S.Yano, J.-C.Peng,andE.Imamovic,SIKA—the multiplexing cold-neutron triple-axis spectrometer atANSTO,Journalof Instrumentation11(10),P10009.
- Lefmannetal. [2006] K.Lefmann, C.Niedermayer, A.Abrahamsen, C.Bahl, N.Christensen, H.Jacobsen, T.Larsen,P.Häfliger, U.Filges,andH.Rønnow,Realizing the full potential of a RITA spectrometer,Physica B: Condensed Matter385-386,1083 (2006).
- Sobolevetal. [2015]O.Sobolev, R.Hoffmann,H.Gibhardt, N.Jünke, A.Knorr, V.Meyer,andG.Eckold,Themultianalyser system of the three axes neutron spectrometer PUMA: Pilotexperiments with the innovative multiplex technique,Nuclear Instruments and Methods in PhysicsResearch Section A: Accelerators, Spectrometers, Detectors and AssociatedEquipment772,63(2015).
- Kikuchietal. [2023]H.Kikuchi, S.Asai,T.J.Sato, T.Nakajima, L.Harriger, I.Zaliznyak,andT.Masuda,A new inelastic neutron spectrometer HODACA,arXiv:2310.11463 [physics.ins-det](2023).
- Lassetal. [2020a]J.Lass, D.Graf, F.Groitl, C.Kägi, R.Müller, R.Bürge, M.Schild, M.S.Lehmann, A.Bollhalder, P.Keller, M.Bartkowiak, U.Filges, F.Herzog, U.Greuter, G.Theidel, L.Testa, V.Favre, H.M.Rønnow,andC.Niedermayer,Design and performance of the multiplexing spectrometer CAMEA,arXiv:2007.14796 [physics.ins-det](2020a).
- Lassetal. [2023]J.Lass, H.Jacobsen,K.M.L.Krighaar,D.Graf, F.Groitl, F.Herzog, M.Yamada, C.Kägi, R.A.Müller, R.Bürge, M.Schild,M.S.Lehmann, A.Bollhalder, P.Keller, M.Bartkowiak, U.Filges, U.Greuter, G.Theidel, H.M.Rønnow, C.Niedermayer,andD.G.Mazzone,Commissioning of thenovel Continuous Angle Multi-energy Analysis spectrometer at the PaulScherrer Institut,Review of Scientific Instruments94,023302 (2023).
- Adibetal. [2011]M.Adib, N.Habib,M.S.El-Mesiry,andM.Fathallah,Characteristics of pyrolytic graphite as a neutronmonochromator,Proceedings of the eighth Nuclear andParticle Physics Conference (NUPPAC-2011),287(2011).
- Groitletal. [2017]F.Groitl, R.Toft-Petersen, D.L.Quintero-Castro, S.Meng, Z.Lu, Z.Huesges, M.D.Le, S.Alimov, T.Wilpert, K.Kiefer, S.Gerischer, A.Bertin,andK.Habicht,MultiFLEXX - the new multi-analyzer at the cold triple-axis spectrometerFLEXX,Scientific Reports7,13637 (2017).
- Limetal. [2015]J.Lim, K.Siemensmeyer,P.Čermák,B.Lake, A.Schneidewind,andD.Inosov,BAMBUS: a new inelastic multiplexed neutron spectrometer forPANDA,Journal of Physics: Conference Series592,012145 (2015).
- Klauszetal. [2021]M.Klausz, K.Kanaki,T.Kittelmann, R.Toft-Petersen, J.O.Birk, M.A.Olsen, P.Zagyvai,andR.J.Hall-Wilton,A simulational study of theindirect-geometry neutron spectrometer BIFROST at the European SpallationSource, from neutron source position to detector position.,J Appl Crystallogr54,263 (2021).
- HBR [2023]HFIR berylliumreflector replacement (2023).
- [19]G.E.Granroth, M.Daum,T.J.Williams, B.Winn, A.A.Aczel, J.Fernandez-Baca, M.Mourigal,andM.D.Lumsden,Incidentbeamline design for a new neutron triple axis spectrometer at the High FluxIsotope Reactor (in preparation).
- Skoulatosetal. [2012] M.Skoulatos, K.Habicht,andK.Lieutenant,Improving energy resolution on neutron monochromator arrays,Journal of Physics: Conference Series340,012019 (2012).
- Birketal. [2014]J.O.Birk, M.Markó,P.G.Freeman, J.Jacobsen, R.L.Hansen, N.B.Christensen, C.Niedermayer, M.Månsson, H.M.Rønnow,andK.Lefmann,Prismatic analyser concept for neutron spectrometers,Review of Scientific Instruments85,113908 (2014).
- Bewley [2019]R.Bewley,FARO: A new type ofneutron spectrometer with flux and resolution optimized,Review of Scientific Instruments90,075106 (2019).
- LefmannandNielsen [1999]K.LefmannandK.Nielsen,McStas, a generalsoftware package for neutron ray-tracing simulations,Neutron News10,20 (1999).
- Salaetal. [2018]G.Sala, J.Y.Y.Lin,V.B.Graves,andG.Ehlers,Conceptual design of CHESS, a new direct-geometryinelastic neutron spectrometer dedicated to studying small samples,Journal of Applied Crystallography51,282 (2018).
- Salaetal. [2022]G.Sala, M.Mourigal,C.Boone, N.P.Butch, A.D.Christianson, O.Delaire, A.J.DeSantis, C.L.Hart, R.P.Hermann, T.Huegle,D.N.Kent, J.Y.Y.Lin, M.D.Lumsden, M.E.Manley, D.G.Quirinale, M.B.Stone,andY.Z,CHESS: The future direct geometry spectrometer at the second targetstation,Review of Scientific Instruments93,065109 (2022).
- Nakajimaetal. [2011]K.Nakajima, S.Ohira-Kawamura, T.Kikuchi, M.Nakamura,R.Kajimoto, Y.Inamura, N.Takahashi, K.Aizawa, K.Suzuya, K.Shibata, etal.,AMATERAS: a cold-neutron disk chopper spectrometer,Journal of the PhysicalSociety of Japan80,SB028 (2011).
- Abernathyetal. [2012]D.L.Abernathy, M.B.Stone, M.Loguillo,M.Lucas, O.Delaire, X.Tang, J.Lin,andB.Fultz,Designand operation of the wide angular-range chopper spectrometer ARCS at theSpallation Neutron Source,Review of Scientific Instruments83 (2012).
- Chaponetal. [2011]L.C.Chapon, P.Manuel,P.G.Radaelli, C.Benson, L.Perrott, S.Ansell, N.J.Rhodes, D.Raspino, D.Duxbury,E.Spill, etal.,Wish: The new powder and singlecrystal magnetic diffractometer on the second target station,Neutron News22,22 (2011).
- Toft-Petersenetal. [2020]R.Toft-Petersen, R.Georgii, M.Schneider,N.Nishiki,andP.Böni,Characterization of pyrolytic graphite with coldneutrons,Nuclear Instruments and Methods in PhysicsResearch Section A: Accelerators, Spectrometers, Detectors and AssociatedEquipment977,164341(2020).
- Lassetal. [2020b]J.Lass, H.Jacobsen,D.G.Mazzone,andK.Lefmann,MJOLNIR: A software package for multiplexingneutron spectrometers,SoftwareX12,100600 (2020b).
- Markó [2014]M.Markó,Analytical calculations for CAMEA,Tech. Rep.(2014).